The physical size of specific detector is known normally, but production business is classified for some sizes that is concerned with the property of detector, such as the well diameter, well depth of detector and dead region. The surface source of even distribution and the sampling method of source particle isotropy sport have been established with the method of Monte Carlo, and gamma ray respond spectral with the 152Eu surface source been calculated. The experiment have been performed under the same conditions. Calculation and experiment results are compared with relative efficiency coincidence method and spectral similar degree coincidence method. According to comparison as a result, detector model is revised repeatedly to determine the calculation model of detector and to calculate efficiency of detector and spectra.