Experimental Research of DT Neutron Sensitivity for Recoil Proton Detection System
Yang GaozhaoNuclear Physics and Chemistry Institute of CAEP,Mianyang,Sichuan Prov,621900)
ABI
Abstract
PIN-recoil proton detection system is designed and constructed for experimental calibration.DT neutron sensitivity of recoil proton detection system is calibrated by recording proton spectrum at K-400 accelerator neutron source.The measured sensitivity well agree with theoretical calculation.The results can furnish experimental indication for design of recoil proton detection system.