MULTIPLE DAMAGE OF THE TIBIAL POLYETHYLENE INSERT IN KNEE JOINT PROSTHESIS (CAN EXCELLENT POSTOPERATIVE OUTCOME BE THE CAUSE OF FRACTURE OF THE TIBIAL INSERT?) CASE REPORT AND LITERATURE REVIEW

Eun Kyoo SongDepartment of Orthopaedic Surgery, Chonnam National University Hwasun Hospital, Seoul, KoreaKamolhuja EshnazarovSaodat AsilovaDepartment of Traumatology, Orthopedics and Neurosurgery, Tashkent Medical AcademyJong Keun SeonDepartment of Orthopaedic Surgery, Chonnam National University Hwasun Hospital, Seoul, Korea
ABI

Abstract

Despite improvements in prosthesis technology and use of high-quality materials in recent years, the number of revisions related to implant failures (4.7%) remains high. Several phenomena were reported in literature as reasons for dislocation and fracture of tibial insert in mobile bearing prosthesis: incorrect positioning of components, discrepancy of extension and flexion balancing or ligament weakness. However, in our cases neither of these causes were observed. The authors consider that bio-physiological and biomechanical aspects of total knee arthroplasty and knee joint prosthesis should be thoroughly studied and implemented into the clinical practice. In the described cases, multiple damage of the insert due to flexion instability and “twist-hyperflexion” in overweight patients is perceived as the reason for failures.