M. HolíkUniversity of West BohemiaF. AhmadovAzerbaijan National Academy of SciencesAzer SadigovNational Nuclear Research CenterG. AhmadovNational Nuclear Research CenterD. BerikovThe Institute of Nuclear PhysicsF. MamedovCzech Technical University in PragueJalal NaghiyevNational Nuclear Research CenterI. ŠteklCzech Technical University in PragueZiraddin Y. SadygovNational Nuclear Research CenterA. G. MammadliAzerbaijan National Academy of SciencesAydan GaribliNational Nuclear Research CenterO. UrbanUniversity of West BohemiaJan ZichUniversity of West Bohemia
Research Squarerepository2022en
ABI

Abstract

Abstract This paper is devoted to gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4x4) with LaBr3(Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved ∼22 % enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3 % while clearly surpassing 4.25 % resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.