Skip to main content
Article

Low-Temperature Degradation of Zirconia and Implications for Biomedical Implants

Jérôme ChevalierMateis, INSA Lyon, CNRS UMR 5510, 69621 Villeurbanne, France;Laurent GrémillardMateis, INSA Lyon, CNRS UMR 5510, 69621 Villeurbanne, France;Sylvain DevilleMaterials Sciences Division, Lawrence Berkeley National Laboratory, Berkeley, California 94720
2007en
ABI

Abstract

This review describes the mechanisms responsible for low-temperature degradation (LTD) of zirconia ceramics and its detrimental consequences for biomedical devices. Special emphasis is given to the critical issue of zirconia degradation actually observed for hip prostheses. Experimental methods to accurately measure and predict LTD in a given zirconia ceramic are presented. Different solutions to inhibit LTD or at least reduce its kinetics are reviewed, with the objective of highlighting alternative options for the generation of new zirconia-based biomedical ceramic devices.

Identifiers

Citations and references

Cited by 20 references