Simulation input spectrum of triple-axis spectrometer and parameter optimization

Song JianmingInstitute of Nuclear Physics and ChemistryChaoqiang HuangYang JianBo ChenChina Academy of Engineering Physics, Mianyang (China). Inst. of Nuclear Physics and Chemistry
ABI

Аннотация

According to the layout and parameters of Cold Neutron Source (CNS) and neutron guides (NG),the parameters of input neutron-beam such as flux,divergence and homogeneity,are calculated by Monte-Carlo simulation at the exit of NG where a neutron triple-axis spectrometer will be built.The results showed the distance(L)between CNS and the entrance of NG should not be less than 0.1m.mo- reover,when the L equal 0.5m the flux increases to the maximal 3.96 10~8cm~(-2) s~(-1).The flux and beam divergence increase slightly as supermirror factor m increasing from 1.5 to 3.0,the vertical maximal di- vergence and flux increase 87.7% and 25% respectively.The vertical and horizontal divergence can be described by a Gaussian function.The horizontal and vertical maximal divergence are 0.85°and 0.81°when m is 1.5,L is 2m,with the flux up to 2.86×10~8 cm~(-2)s~(-1).We give the optimum neutron guides pa- rameter for the TAS.