Dose–response of EBT3 radiochromic films to proton and carbon ion clinical beams

Roberta CastriconiUniversità di Napoli Federico IIMario CioccaCentro Nazionale di Adroterapia OncologicaAlfredo MirandolaCentro Nazionale di Adroterapia OncologicaC. SiniOspedale San Raffaele di MilanoS. BroggiOspedale San Raffaele di MilanoMarco SchwarzNational Institute of Nuclear Physics (INFN)Francesco FracchiollaCentro di Protonterapia di TrentoM. MartišíkováHeidelberg Institute for Radiation Oncology (HIRO)Giulia AricòDepartment of Radiation Oncology and Radiation Therapy, Heidelberg University Hospital, Heidelberg, GermanyGiovanni MettivierDipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, and INFN Sezione di Napoli, Napoli, ItalyPaolo RussoUniversità di Napoli Federico II
2016en
ABI

Annotatsiya

We investigated the dose-response of the external beam therapy 3 (EBT3) films for proton and carbon ion clinical beams, in comparison with conventional radiotherapy beams; we also measured the film response along the energy deposition-curve in water. We performed measurements at three hadrontherapy centres by delivering monoenergetic pencil beams (protons: 63-230 MeV; carbon ions: 115-400 MeV/u), at 0.4-20 Gy dose to water, in the plateau of the depth-dose curve. We also irradiated the films to clinical MV-photon and electron beams. We placed the EBT3 films in water along the whole depth-dose curve for 148.8 MeV protons and 398.9 MeV/u carbon ions, in comparison with measurements provided by a plane-parallel ionization chamber. For protons, the response of EBT3 in the plateau of the depth-dose curve is not different from that of photons, within experimental uncertainties. For carbon ions, we observed an energy dependent under-response of EBT3 film, from 16% to 29% with respect to photon beams. Moreover, we observed an under-response in the Bragg peak region of about 10% for 148.8 MeV protons and of about 42% for 398.9 MeV/u carbon ions. For proton and carbon ion clinical beams, an under-response occurs at the Bragg peak. For carbon ions, we also observed an under-response of the EBT3 in the plateau of the depth-dose curve. This effect is the highest at the lowest initial energy of the clinical beams, a phenomenon related to the corresponding higher LET in the film sensitive layer. This behavior should be properly modeled when using EBT3 films for accurate 3D dosimetry.

Hali tarjima qilinmagan

Identifikatorlar

Iqtiboslar va manbalar