Improved evaluation of measurement uncertainty and traceability in vibrometer calibration systems
Litsenziya: cc-by
Annotatsiya
This study presents an improved approach for evaluating measurement uncertainty and ensuring metrological traceability in vibrometer calibration systems. The proposed method enhances the implementation of ISO 16063-21 and GUM guidelines by integrating a detailed uncertainty budget with frequency-dependent correction factors. The calibration experiments were performed in the range of 10-1000 Hz using a reference comparison technique. The expanded uncertainty of the developed system was estimated as <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mi>U</mml:mi> <mml:mfenced separators="|"> <mml:mrow> <mml:mi>k</mml:mi> <mml:mo>=</mml:mo> <mml:mn>2</mml:mn> </mml:mrow> </mml:mfenced> <mml:mo>=</mml:mo> <mml:mi mathvariant="normal"> </mml:mi> </mml:math> (1.7-2.1) %, depending on frequency and environmental stability. The novelty of the proposed approach lies in the refined identification of dominant uncertainty components – including reference sensor calibration, vibration table stability, and signal processing repeatability – with quantitative contribution analysis. The obtained results confirm the method’s capability to improve traceability and reproducibility in national and industrial calibration laboratories. The study provides a practical framework for enhancing accuracy, comparability, and long-term reliability of vibration measurement systems.
Hali tarjima qilinmagan