Asosiy kontentga oʻtish
Maqola

Real‐time reaction monitoring by probe electrospray ionization mass spectrometry

Zhan YuCollege of Chemistry and Biology Shenyang Normal University Shenyang ChinaLee Chuin ChenClean Energy Research Center, University of Yamanashi, Kofu, JapanRosa Erra‐BalsellsFundación Ciencias Exactas y NaturalesHiroshi NonamiPlant Biophysics/Biochemistry Research Laboratory, Faculty of Agriculture, Ehime University, Matsuyama, JapanKenzo HiraokaClean Energy Research Center, University of Yamanashi, Kofu, Japan
2010en
ABI

Annotatsiya

Probe electrospray ionization (PESI) is a modified version of the electrospray ionization (ESI), where the capillary for sampling and spraying is replaced by a solid needle. High tolerance to salts and direct ambient sampling are major advantages of PESI compared with conventional ESI. In this study, PESI-MS was used to monitor some biological and chemical reactions in real-time, such as acid-induced protein denaturation, hydrogen/deuterium exchange (HDX) of peptides, and Schiff base formation. By using PESI-MS, time-resolved mass spectra and ion chromatograms can be obtained reproducibly. Real-time PESI-MS monitoring can give direct and detailed information on each chemical species taking part in reactions, and this is valuable for a better understanding of the whole reaction process and for the optimization of reaction parameters. PESI-MS can be considered as a potential tool for real-time reaction monitoring due to its simplicity in instrumental setup, direct sampling with minimum sample preparation and low sample consumption.

Hali tarjima qilinmagan

Identifikatorlar

Iqtiboslar va manbalar

2 ta iqtibos0 ta foydalanilgan manba