LiF detectors application for detection of X-ray radiation from plasma

Yu. L. Bakshaev Russian Research Center, Kurchatov Institute, 123182, Kurchatov sq. 1, Moscow, Russia; Phone: 095 1969509 Fax: 095 1969509 E-mail: [email protected]
1999en
ABI

Annotatsiya

Abstract The LiF detectors of a commercial device with the measurement range from 25 keV to 1.3 MeV were successfully used for of the monitoring X-ray from the imploding liner plasma on the high current pulse generator S-300. The detector readings were discovered to decrease more than ten times behind a thin lavsan filter of 2 μm thick in comparison with the open detector readings. This result demonstrates a possibility of using the LiF detector for registration of soft X-ray less than 1 keV. The detectors work in the regime of fuli absorption of the incident X-ray radiation up to the energy of about 10 keV. The stability of the detector readings under identical experimental conditions allows one to use them for integral monitoring of the X-ray plasma radiation. Key Words: Thermoluminescent LiF detectorsoft X-ray radiationliner plasma

Hali tarjima qilinmagan

Identifikatorlar

Iqtiboslar va manbalar

2 ta iqtibos0 ta foydalanilgan manba