Determining in vivo biomechanical properties of the cornea with an ocular response analyzer

David A. LuceReichert Inc., Depew, New York, USA. [email protected]
2005en
ABI

Annotatsiya

PURPOSE: To study the results of an ocular response analyzer (ORA) to determine the biomechanical properties of the cornea and their relationship to intraocular pressure (IOP). SETTING: Reichert Inc., Depew, New York, USA. METHODS: The ORA (Reichert) makes 2 essentially instantaneous applanation measurements that permit determination of corneal and IOP effects. RESULTS: Measurements of several populations indicate that corneal hysteresis, a biomechanical measure, varied over a dynamic range of 1.8 to 14.6 mm Hg and was only weakly correlated with corneal thickness (r(2)=0.12); this is related to the observation that some subjects with relatively thick corneas have less-than-average corneal hysteresis. Corneal hysteresis changes diurnally, presumably as a result of hydration changes. Keratoconus, Fuchs' dystrophy, and post-LASIK patients demonstrated low corneal hysteresis. CONCLUSION: The corneal hysteresis biomechanical measure may prove valuable for qualification and predictions of outcomes of refractive surgery and in other cases in which corneal biomechanics are important.